Writer: drmarthahall
drmarthahall
12 minutes ago
5 min read

Short answer: FDA treats the wearable in your trial as part of your evidence, not just a way to collect it. Its final guidance on digital health technologies (DHTs) asks you to show the device is fit for purpose through verification, validation and usability, and it names comfort, materials and skin contact along the way. If participants can't wear it correctly for the full study, your endpoint is at risk.


What does FDA's DHT guidance actually ask for?

FDA finalized Digital Health Technologies for Remote Data Acquisition in Clinical Investigations in December 2023. CDER led it with CBER, CDRH and the Oncology Center of Excellence, and as of November 2026 it is still the agency's final word on the subject.

The core idea is "fit for purpose." Sponsors should show three things:

  • Verification: the sensor measures its physical parameter, such as acceleration or temperature, accurately and precisely.

  • Validation: the DHT appropriately assesses the clinical event or characteristic in the proposed participant population.

  • Usability: participants can use the DHT as intended, at home, without a coordinator standing next to them.


These apply whether or not your DHT is a regulated medical device. Most commentary focuses on software, data flow and source records. I came to wearables through apparel design and biomechanics, so I read the guidance from the skin out. The body shows up far more often than most sponsors notice.


Where does the guidance talk about the body?

Read it closely and you will find a wearability specification tucked inside a data guidance:

  • Design: it lists "material, size, weight, appearance, portability" as features that can influence use, and says "comfort and convenience" may affect a participant's ability and willingness to use a wearable.

  • Population: it asks sponsors to consider "education, language, age, health/physical condition, and technical aptitude" when choosing a DHT.

  • Usability: "Usability evaluations should assess the ability for participants to efficiently use the DHT in a remote setting."

  • Clinical risk: "The physical features of the DHT should be evaluated for discomfort" and risk of injury, including "skin contacting components" that cause irritation.

  • Measurement variation: an appendix example names "body morphology, skin color" and variation in sensor placement as sources of variability.

  • Missing data: sponsors should have a "plan in place to reduce the potential for missing data."

Every one of those lands on the physical device. None of them can be solved by firmware alone.


What has changed in 2026?

The 2023 guidance hasn't been revised, but FDA's attention to DHTs has picked up sharply this year:

  • March 2026: FDA published a request for information on advancing DHTs in drug and biologic trials (docket FDA-2026-N-2476). Comments closed June 1 and will inform future guidance.

  • July 2026: FDA opened a $1.1 million research funding opportunity (RFA-FD-26-012) for DHT measurement work in drug development.

  • August 2026: CBER, CDER, CDRH and the Oncology Center of Excellence jointly published a white paper, Key Considerations for the Development and Use of Digitally Derived Measures for Clinical Investigations.


That white paper is not guidance, but it is the clearest signal yet of where reviewers' questions are heading. For DHTs worn for prolonged or repeated use, it says sponsors "should evaluate adherence, user burden, sustained usability" and potential user fatigue across the study. It adds that "the tolerability of the DHT for the target population" should be taken into account, and it flags "incorrect placement of a wearable DHT (e.g., wrist versus hip)" as a source of measurement error. FDA's Digital Health Center of Excellence summed up the theme: start with what matters to patients, not with the sensor.

In plain terms, the bar is moving from "usable on day one" to "still worn correctly on day 60."


Why is wearability a verification and validation question?

This is the part I want clinical operations and digital endpoint leads to sit with.


Verification assumes the device stays put. Bench accuracy depends on consistent placement and skin contact. On a real body, straps get loosened for comfort, housings rotate on a thin wrist, and adhesives lift with sweat. The sensor is still accurate. It is just no longer measuring what you verified.


Validation is tied to your population. FDA asks whether the DHT works "in the proposed participant population." If your inclusion criteria include edema, tremor, neuropathy, fragile skin or a wide range of body sizes and skin tones, a device validated on healthy volunteers hasn't been validated for your trial.


Usability includes putting it on. A clasp that someone with limited hand strength can't close is a use error. So is a band worn upside down because both ends look the same. These are the participants many trials most need to study.

Missing data starts at the skin. Reminders, alerts and coordinator outreach all help, and the guidance points to measures like them. But they treat the symptom. The cheapest place to reduce missing data is a device participants forget they're wearing.


How do you build wearability into your DHT evidence?

Here is the approach I use with sponsors before a protocol locks:

  1. Write wearability requirements next to performance specs. I use the 5F Framework: Fit, Function, Fabric, Fastener and Fashion. Make each one testable, such as "stays in position through sleep," "put on and taken off alone in under a minute" or "no skin irritation across the planned wear period."

  2. Verify on bodies, not only benches. Check signal quality with the device worn the way participants will actually wear it, including loosened straps, sweat, sleep and placement drift.

  3. Validate in the population you'll enroll. Match age, body size, skin tone, dexterity and condition-specific changes like swelling to your inclusion criteria.

  4. Run a short at-home wear study for the full planned duration. Record when and why the device comes off. That one study gives you usability evidence, a tolerability read and real numbers for your missing-data plan.

  5. Treat the body interface as part of the DHT. The guidance warns that significant measurement changes after updates "may invalidate the results." I apply the same discipline to a new strap, band size or adhesive: record it and confirm the device is still fit for purpose.


Key takeaways

  • FDA's final DHT guidance names material, size, weight, comfort and skin contact. Wearability is already in scope.

  • "Fit for purpose" means fit for the bodies in your trial, not a healthy volunteer.

  • FDA's August 2026 white paper adds adherence, user burden, sustained usability and tolerability to the conversation.

  • The most effective missing-data strategy is a device people keep on.


FAQ

Does the DHT guidance apply to wearables that aren't FDA-cleared? Yes. Verification and validation apply whether or not the DHT meets the definition of a device; sponsors still need to show it is fit for purpose.


Does FDA require a usability study? The guidance says usability evaluations should be conducted to show the DHT can be used as intended. How you do that depends on the device, the population and the context of use.


Is the August 2026 white paper binding? No. It is a discussion paper that summarizes considerations from existing FDA guidance. It is still worth reading, because it shows what reviewers are likely to ask.